SOMATOM Definition EdgeExceeding expectations

Confronted with increasingly complex clinical requirements and rising numbers of patients, medical institutions are expected to perform at the limits of capacity every day. The new SOMATOM Definition Edge helps you not only meet, but exceed these expectations by improving your institution’s process efficiency and patient outcome in all clinical capabilities – from contrast media-efficient TAVI planning to precise therapy response management, from low dose therapy control to optimized emergency care workflow.

  • Exceeding expectations in:
  • Cardiology
  • Emergency Medicine
  • Oncology

Features & Benefits


CT Scanner - SOMATOM Definition Edge - Cardiology

Making highly precise plaque differentiation part of your clinical routine will not only improve patient outcome, but also increase process efficiency, as unmistakingly clear findings will drastically reduce the number of discussions about the state of the plaque. Covering greater volumes faster, improving contrast media efficiency in low kV TAVI planning, and introducing reliable, high-speed triple rule-out scanning will expand your institution’s clinical capabilities – and help you exceed the expectations that come with increasingly complex cardiac cases.
How is this possible? Learn more about the new Stellar detector and the STRATON tube.


CT Scanner - SOMATOM Definition Edge - Emergency Medicine

Emergency scanning is as much about time as it is about image quality.
You will be able to optimize process efficiency with solutions that let you not only improve emergency workflow but also substantially reduce door-to-image time, from pediatric to obese patients. The new SOMATOM Definition Edge enables excellent tissue evaluation, allows for the scanning of pediatric and obese patients without dose discussions, and facilitates a minimized sedation and breath hold with a pitch of 1.7 at 23 cm/sec, so you can reduce door-to-image and optimize your overall emergency workflow. Learn more about the Stellar detector and the STRATON tube, our complete FAST CARE technology, Adaptive 4D Spiral Plus, TrueSignal Technology, and iMAR2, 3.


CT Scanner - SOMATOM Definition Edge - Oncology

With around a quarter of therapies adjusted after response assessment, a key challenge in oncology is finding out if the tumor is responsive to therapy. Making therapy response assessment faster, more reliable, and easier to manage, will directly benefit your workforce as well as your patients. Expanding your clinical capabilities with improved, low dose therapy control, and CT-based early tumor identification will benefit your institution both in daily practice and in the long run: the more reliable and dose-efficient the technology gets, the greater its role will become for preventive care, as well.
Find out how ADMIRE4, Adaptive 3D Intervention Suite, TwinBeam Dual Energy2 and True Dual Energy CT Applications make it happen.

Clinical Use


Introducing a rotation speed of 0.28 s, and a temporal resolution of 142 ms to clinical CT routine, the new SOMATOM Definition Edge provides excellent image quality in cardiac imaging even for higher heart rates of more than 80 bpm.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 100 kV, eff. 98 mAs

Scan time: 3.7 s

DLP: 179 mGy cm

Scan length: 137 mm

CTDIvol: 11.39 mGy

Rotation time: 0.28 s

Eff. dose: 2.5 mSv

Heart rate: 81 bpm

 


With an acquisition speed of up to 23 cm/s (pitch 1.7), the new SOMATOM Definition Edge boosts contrast media efficiency and substantially shortens scan times, especially in pre-procedural TAVI/TAVR planning.

CT Scanner - SOMATOM Definition Flash - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 100 kV, 70 eff. mAs

Pitch: 1.7

DLP: 172 mGy cm

Scan time: 2.5 s

CTDIvol: 2.77 mGy

Scan length: 580 mm

Eff. dose: 2.49 mSv

Rotation time: 0.28 s

Heart rate: 77 bpm


Routinely high rotation speed, combined with Edge technology and 0.5 mm slices, delivers additional diagnostic information for more precise plaque differentiation.

CT Scanner - SOMATOM Definition Flash - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 80 kV, 75 eff. mAs

Flash scan mode

DLP: 72.65 mGy cm

Scan time: 6.0 s

CTDIvol: 4.38 mGy

Scan length: 146 mm

Eff. dose: 1.0 mSv

Rotation time: 0.28 s

Heart rate 67 bpm


The Adaptive 4D Spiral puts perfusion scanning on a new level. Being able to evaluate perfusion deficits in a very short time, and to effortlessly acquire scan range beyond detector coverage, greatly improves stroke assessment and workflow.

CT Scanner - SOMATOM Definition Flash - Clinical Case

VPCT

CTA

Collimation: 32 x 1.2 mm

Collimation: 128 x 0.6 mm

Scan time: 44.0 s

Scan time: 6.0 s

Scan length: 90 mm

Scan length: 370 mm

Rotation time: 0.5 s

Rotation time: 0.5 s

Tube settings: 80 kV, 200 eff. mAs

Tube settings: 120 kV, 211 eff. mAs

DLP: 2.8 mGy cm

DLP: 724 mGy cm

CTDIvol: 237.49 mGy

CTDIvol: 18.47 mGy

Eff. dose: 5.88 mSv

Eff. dose: 3.04 mSv


The SOMATOM Definition Edge provides outstanding tube and generator power and faster scan speeds (up to 23 cm/s) to guarantee optimal image quality even in challenging cases, for example obese patients.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 120 kV, 130 eff. mAs

Pitch: 1.7

DLP: 260 mGy cm

Scan time: 1.1 s

CTDIvol: 8.83 mGy

Scan length: 252 mm

Eff. dose: 3.64 mSv

Rotation time: 0.28 s

 


In challenging cases, such as this 9 months old baby with suspected bronchia stenosis, the SOMATOM Definition Edge’s faster pitch (of 1.7) and higher rotation time (of 0.28s) allow for acquiring diagnostic information with free breathing at an ultra low dose of DLP 2 mGy cm.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 100 kV, 4 eff. mAs

Scan time: 0.6 s

DLP: 2 mGy cm

Scan length: 133 mm

CTDIvol: 0.14 mGy

Rotation time: 0.28 s

Eff. dose: 0.17 mSv


iMAR2,3 technology applies iterative corrections to reduce metal artifacts effectively. Even in challenging cases, such as two hip prostheses, the pelvic anatomical structures can be clearly visualized.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 100 kV, 372 eff. mAs

Scan time: 9.5 s

DLP: 682.4 mGy cm

Scan length: 448.5 mm

CTDIvol: 14.7 mGy

Rotation time: 0.5 s

Eff. dose: 10.2 mSv


CT angiography acquired with TwinBeam Dual Energy, using automatic bone removal, converts a large dataset into three dimensional visualizations very conveniently and easily.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 64 x 0.6 mm

Tube settings: AuSn120 kV, 301 eff. mAs

Scan time: 44 s

DLP: 1,007.8 mGy cm

Scan length: 1,534 mm

CTDIvol: 6.44 mGy

Rotation time: 0.33 s

Eff. dose: 8 mSv


ADMIRE4, the latest generation of iterative reconstruction, greatly enhances image quality while reducing radiation dose. In this pediatric cerebral examination, great grey-white matter differentiation was achieved using a radiation dose of CTDIvol 21.1 mGy in combination with ADMIRE.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 120 kV, 146 eff. mAs

Scan time: 5.7 s

DLP: 329.9 mGy cm

Scan length: 140 mm

CTDIvol: 21.1 mGy

Rotation time: 1 s

Eff. dose: 1.3 mSv


FAST Spine helps streamline workflows when reconstructing spine datasets. In challenging cases, such as this 13 years old boy with scoliosis, the benefits of reduced door-to-image time boost diagnostic accuracy and confidence – and make a huge difference in patient management.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 100 kV, 52 eff. mAs

Scan time: 18 s

DLP: 96 mGy cm

Scan length: 541 mm

CTDIvol: 1.64 mGy

Rotation time: 1.0 s

Eff. dose: 1.44 mSv


The Stellar Detector technology, in combination with high-end iterative reconstruction algorithms, such as SAFIRE5 and ADMIRE, facilitates ruling out pulmonary disease at an early stage, using ultra low doses.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 100 kV, 10 eff. mAs

Scan time: 1.9 s

DLP: 15 mGy cm

Scan length: 300 mm

CTDIvol: 0.39 mGy

Rotation time: 0.28 s

Eff. dose: 0.21 mSv


Metal artifacts can impact the visualization of the surrounding tissues, as shown in this case – a small tumor, adjacent to a tibial metal fixation, therefore needs to be corrected properly. With iMAR2,3, Siemens presents an exclusive iterative metal artifact reduction technique that enhances diagnostic confidence and provides clear guidance for biopsy.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 128 x 0.6 mm

Tube settings: 120 kV, 157 eff. mAs

Scan time: 4.4 s

DLP: 173.3 mGy cm

Scan length: 147 mm

CTDIvol: 10.61 mGy

Rotation time: 0.5 s

Eff. dose: 2.6 mSv


A follow-up scan of a liver metastasis after microwave ablation treatment acquired by TwinBeam Dual Energy2 shows a hypodense lesion in the liver and in the right kidney. The image contrast is greatly increased by monoenergetic imaging. And both lesions show no significant enhancement in the fused VNC/ iodine image. Therefore, residual tumor tissue can be confidently excluded and a kidney cyst can be diagnosed.

CT Scanner - SOMATOM Definition Edge - Clinical Case

Collimation: 64 x 0.6 mm

Tube settings: AuSn120 kV, 410 eff. mAs

Scan time: 19.6 s

DLP: 528.1 mGy cm

Scan length: 634 mm

CTDIvol: 8.8 mGy

Rotation time: 0.28 s

Eff. dose: 8.7 mSv

View Customer Outcome Stories in the download section.

 

Technical Specifications

Detector

Stellar detector

Number of acquired slices

128

Number of reconstructed slices

384

Spatial resolution

0.30 mm

Rotation time

0.28 s6

Temporal resolution

142 ms

Generator power

100 kW6

kV steps

70, 80, 100, 120, 140 kV

Max. scan speed

230 mm/s

Table load

up to 307 kg / 676 lbs6

Gantry opening

78 cm


CT Scanner - SOMATOM Definition Edge - Stellar detector

Lower signal, lower voltage, lower dose – the new Stellar detector boosts both patient outcome and risk reduction without any compromise in image quality and diagnostic reliability. Based on full electronic integration, Siemens’ exclusive TrueSignal technology significantly reduces noise and cross-talk. Instead of increasing dose, it makes more efficient use of the initially available quants per voxel. In addition, the Stellar detector provides a homogeneous slice profile with the same spatial resolution and SNR in the entire Field of View. And with no more need to switch bandwidth, it also benefits your patients with an extended dynamic range (HiDynamics), as the detector’s sensitivity for visualizing finer structures, especially for low kV datasets, is substantially increased.


CT Scanner - SOMATOM Definition Edge

Introducing dose-efficient, comprehensive Dual Energy capabilities to virtually all patients, the new SOMATOM Definition Edge offers greater versatility than any other single source CT system. TwinBeam Dual Energy2 allows simultaneous acquisition of high and low kV datasets in a single CT scan, enabling high-contrast dynamic applications. To create two X-ray spectra simultaneously from one tube, the STRATON tube assembly system generates a pre-filtered X-ray beam, creating two different x-ray spectra before it reaches the patient. With all syngo.via applications now available for all single source DE cases, image acquisition is possible for all rotation times (up to 0.28 seconds), for the full field-of-view of 50 cm, and with the full number of projections for both spectra.


CT Scanner - SOMATOM Definition Edge - ADMIRE

Innovation leader Siemens has been at the forefront of radiation reduction for decades – scans at sub-mSv doses have found their way into clinical routine. However, there isn’t one dose level that fits everyone. Every clinical question and every single patient require individual, meticulous dose management. Siemens’ comprehensive Combined Applications to Reduce Exposure (CARE), in combination with the STRATON tube and iterative reconstruction methods SAFIRE5 and ADMIRE4, enable low kV imaging down to 70 kV, enhance contrast media for advanced diagnostic information, and allow for substantial dose savings. In addition, the Adaptive Dose Shield eliminates overradiation pre- and post-spiral to the patient by dynamically moving shields into
place on the X-ray tube – blocking clinically irrelevant dose in every single spiral acquisition.

CT Scanner - SOMATOM Definition Edge - Stellar detector

Lower signal, lower voltage, lower dose – the new Stellar detector boosts both patient outcome and risk reduction without any compromise in image quality and diagnostic reliability. Based on full electronic integration, Siemens’ exclusive TrueSignal technology significantly reduces noise and cross-talk. Instead of increasing dose, it makes more efficient use of the initially available quants per voxel. In addition, the Stellar detector provides a homogeneous slice profile with the same spatial resolution and SNR in the entire Field of View. And with no more need to switch bandwidth, it also benefits your patients with an extended dynamic range (HiDynamics), as the detector’s sensitivity for visualizing finer structures, especially for low kV datasets, is substantially increased.

CT Scanner - SOMATOM Definition Edge - STRATON tube with z-Sharp

The STRATON™ tube with z-Sharp™ and its revolutionary design based on a direct anode cooling eliminate the need for heat storage and results in an unmatched compact design, allowing true temporal resolution of up to 150 ms, and a high isotropic resolution of 0.33 mm at any scan and rotation speed, and at any position within the scan field. This, for instance, allows to significantly reduce motion artifacts of the heart to perform accurate stenosis measurements or stent planning with outstanding precision. It facilitates fast whole body sub-mm imaging at highest pitch revealing outstanding details in long-range vascular studies or polytrauma patients.

CT Scanner - SOMATOM Definition Edge - SAFIRE

With SAFIRE Siemens has set the benchmark on low dose imaging with raw-data based iterative reconstruction. For the first time, raw data (or sinogram data) are actually being utilized in the iterative image improvement process. With this, SAFIRE can achieve a radiation dose reduction of up to 60%5. Or instead improve image quality (contrast, sharpness and noise).

SAFIRE in 3 steps:

  1. Multiple iterative loops performed in the raw-data domain correct geometrical imperfections and other artifacts
  2. Image reconstruction from the “corrected” raw-data
  3. Multiple iterative loops performed in the image domain remove image noise for a final and excellent image quality
CT Scanner - SOMATOM Definition Edge - ADMIRE

The lower the dose the more difficult it is to reconstruct images with an adequate image quality. This rule is common in CT. Now with Advanced Modeled Iterative Reconstruction ADMIRE, the newest generation in iterative reconstruction, clinicians and radiologists have access to enhanced image quality with a natural image impression and all clinical applications at a dose level that exceed all expectations. ADMIRE makes integrating iterative reconstruction into daily routine easy: Thick slices are reconstructed at a natural image impression, even from ultra-low dose scans. The reconstructed datasets can easily be stored in PACS or on film.

The new benchmark in single source dual energy – TwinBeam Dual Energy – allows for simultaneous acquisition of high and low kV datasets in a single CT scan, enabling high-contrast dynamic applications. Making dose-efficient, comprehensive Dual Energy capabilities routinely applicable to virtually all patients without compromising on image quality, the new SOMATOM Definition Edge offers greater versatility than any other single source CT system.

CT Scanner - SOMATOM Definition Edge - FAST CARE

The FAST CARE technology suite exceeds expectations in both of CT’s essential dimensions: Getting reliable results with as little radiation exposure as possible, and getting them fast.

FAST applications – our Fully Assisting Scanner Technologies –, make time-consuming and complex procedures faster and more intuitive. Scans will be more reproducible and less prone to errors. FAST Planning, for example, features immediate, organ-based setting of scan and recon ranges with no manual adjustments.

CARE technologies – our Combined Applications to Reduce Exposure –, offer excellent image quality at the right dose level, making CT scanning safer even for the most delicate patients, especially children. The CARE Dose Configurator, for example, allows you to better manage dosage with respect to the patient's anatomy when adjusting scan parameters.

CT Scanner - SOMATOM Definition Edge - Adaptive 4D spiral

The SOMATOM Definition Edge features Adaptive 4D Spiral Plus for long-range dynamic imaging of up to 48 cm. It exceeds expectations regarding the visualization and evaluation of the entire thoracic aorta in separate phases by capturing the arterial and venous phase in a single scan. So you’re able to not only delineate aneurysms and other vascular diseases, but also to identify the source of bleedings or assess venous thrombosis. It provides full organ coverage in 4D, be it stroke assessment of the brain or body perfusion studies – intelligent filtering with 4D Noise Reduction allows acquisition with up to 50 % lower dose.

CT Scanner - SOMATOM Definition Edge - True Signal

The lower the signal, the higher the noise, the poorer the image quality?
Not true anymore. Based on the full electronic integration of the Stellar detector, TrueSignal technology significantly reduces electronic noise, especially in low kV images. The resulting signal-to-noise ratio is improved manyfold, and low-signal datasets benefit from increased sharpness and clarity, as the detector can now differentiate much better between the signal of individual voxels and the surrounding image data.

CT Scanner - SOMATOM Definition Edge - iMAR

Metal fillings and implants cause artifacts that often result in undiagnostic image quality, costing time for planning and potentially threatening therapy outcomes. iMAR2,3 is the industry’s most effective algorhythm focused on the removal of metal artifacts for enhanced diagnostic image quality. The minimal introduction of new interpolation artifacts, and the preservation of natural image impression and edge information will help streamline workflows and improve diagnostic confidence – with iMAR, surgical, prosthetic and dental fillings as well as pacemakers are not hiding insights anymore.

1
2
3
4
5
6