CABG of the right gastroepiploic artery and of the left internal mammary artery

Shaodong Cao, MD1; Bo Han, MD2; Xi Zhao, MD2; Tong Zhang, MD1

1 Department of Radiology, The No. 4 Hospital of Harbin Medical University, Jilin, P. R. China

 2 Siemens Healthineers, China

21.04.2023

A 42-year-old male patient, suffering from exertional chest tightness and shortness of breath for the past 4 days, presented himself to the hospital. His symptoms could be relieved after a few minutes of rest, had been however progressive. A coronary angiography was performed which revealed severe stenoses in the mid right coronary artery (RCA) and the proximal left anterior descending artery (LAD). Coronary artery bypass grafting (CABG) was immediately performed. Seven days later, a CT angiography (CTA) was requested to assess the patency of the grafts as well as of the anastomoses.

The CTA images showed two grafts – the right gastroepiploic artery (RGEA), arising from the gastroduodenal artery (GDA), was anastomosed to the distal RCA. The left internal mammary artery (LIMA), originating from the left subclavian artery (LSA), was anastomosed to the mid LAD. Both grafts were patent without kinking or twisting. The anastomoses were also patent. Mixed plaques in the mid RCA and the proximal LAD, causing severe stenoses, were seen and were consistent with the coronary angiography findings. A calcified plaque in the proximal circumflex artery (Cx) was visualized, causing no significant stenosis. The patient recovered uneventfully and was discharged.

A cinematic VRT image and two curved MIP images show an overview of the two grafts – the RGEA, arising from the GDA, is anastomosed to the distal RCA; the LIMA, originating from the LSA, is anastomosed to the mid LAD. Both grafts and anastomoses are patent.

Courtesy of Department of Radiology, The No. 4 Hospital of Harbin Medical University, Jilin, P. R. China

Figs. 1: A cinematic VRT image (Fig. 1a) and two curved MIP images (Figs. 1b & 1c) show an overview of the two grafts – the RGEA, arising from the GDA, is anastomosed to the distal RCA (arrows); the LIMA, originating from the LSA, is anastomosed to the mid LAD (dashed arrows). Both grafts and anastomoses are patent.

Curved MPR and MIP images show mixed plaques in the mid RCA and the proximal LAD, causing severe stenoses. A calcified plaque in the proximal Cx is seen, causing no significant stenosis. Both anastomoses to the LAD and to the RCA are patent.

Courtesy of Department of Radiology, The No. 4 Hospital of Harbin Medical University, Jilin, P. R. China

Fig. 2: Curved MPR (Figs. 2b & 2d) and MIP images (Figs. 2a & 2c) show mixed plaques in the mid RCA and the proximal LAD, causing severe stenoses. A calcified plaque in the proximal Cx is seen, causing no significant stenosis. Both anastomoses to the LAD (dashed arrow) and to the RCA (arrow) are patent.

CABG surgery is performed to bypass the diseased coronary arteries to maintain blood supply to the affected areas of the myocardium. The choice of conduit has been a major concern for cardiac surgeons desiring a good surgical outcome. The RGEA is most suitable for grafting the distal right coronary artery and the posterior descending artery since this site is the nearest for the in-situ GEA graft. According to American Heart Association guidelines, an RCA with more than a 90% stenosis is considered as a proper target for applying a GEA graft. [1] The LIMA is most commonly used as an in-situ graft to revascularize the LAD, due to its proximity to the LAD and favorable patency rates. [2] As the long-term clinical outcome depends upon the patency of the CABG, proper follow-ups are essential. Coronary CTA is a non-invasive imaging method and is capable of depicting the entire course of the grafts, providing detailed information for the evaluation of the vessel lumen and the anastomoses in three dimensions. To overcome the technical challenges, such as the presence of calcifications and clip material over a long scan range obscuring the visualization of the lumen, as well as elevated, irregular heart rates of the patients causing motion artifacts, sophisticated hardware and intelligent software are required and have been developed. This case is performed on a Dual Source CT scanner, SOMATOM Force, which provides a high temporal resolution of approximately 66 ms – a feature that is essential for motion-free image acquisition. A retrospective ECG gated spiral mode is available for scanning patients with elevated, irregular heart rates. A lower kV setting of 90 kV is selected by the system using CARE kV – an automated feature that adjusts the tube voltage, tailored to the individual patient, the system capabilities and the clinical task, improving the contrast enhancement and reducing the amount of contrast agent needed. These technical advancements support physicians for post CABG evaluation.

Scanner

Scan area

Chest/Abdomen

Scan mode

Retrospective ECG
gated spiral mode

Scan length

350 mm

Scan direction

Cranio-caudal

Scan time

4.8 s

Tube voltage

90 kV

Effective mAs

351 mAs

Dose modulation

CARE Dose4D

CTDIvol

24.8 mGy

DLP

953.8 mGy*cm

Rotation time

0.25 s

Pitch

0.32

Slice collimation

192 x 0.6 mm

Slice width

0.75 mm

Reconstruction increment

0.4 mm

Reconstruction kernel

Br40

Heart rate

85 – 89 bpm

Contrast

350 mg/mL

Volume

64 mL + 38 mL saline

Flow rate

4.2 mL/s

Start delay

Bolus tracking triggered at 100 HU in the descending aorta + 6 s