Atellica® 1500 Automated Urinalysis System
A urinalysis system so advanced, it lets you do less and see more.

Combining the CLINITEK Novus® Analyzer and the new Atellica® UAS 800 Analyzer* into one completely automated unit, the Atellica 1500 Automated Urinalysis System* sets the new standard for accuracy and efficiency. It is a truly digital automated urinalysis system, and lets you manage more samples with less staff in shorter time, while never compromising on high-quality results.
- Combines 2 advanced analyzers, the trusted CLINITEK Novus Analyzer, and the new Atellica UAS 800 Analyzer
- The CLINITEK Novus Analyzer combines trusted urine strip technology with cutting-edge digital imaging to deliver consistent, high-quality results
- The new Atellica UAS 800 Analyzer is a completely automated urine sediment analyzer designed to minimize the need for manual microscopy
- Urine chemistry and sediment technologies streamlines detection of UTIs and early kidney disease
Features & Benefits

Do less and see more with a truly digital automated urinalysis system that sets the new standard for accuracy and efficiency.

The System
Atellica 1500 Automated Urinalysis System
The Atellica® 1500 Automated Urinalysis System* lets you manage more samples with less staff in a shorter time, while never compromising on high-quality results.
- Measurement technology: Digital microscopy imaging with a dual-focusing mechanism to generate clear images
- Combines dry-pad urine chemistry and powerful digital microscopy into a single, streamlined unit
- Throughput range: Up to 240 samples per hour
- Streamlines the detection of UTIs and early kidney disease for better clinical outcomes

The Analyzers
CLINITEK Novus® Automated Urine Chemistry Analyzer
- Digital Camera Module: LED and colored CCD camera that captures full RGB color spectrum to calculate color changes for each parameter, instead of a 3-4 fixed wavelength
- Optical Sensors: Refractometry method that calculates clarity and specify gravity

Atellica® UAS 800 Urine Sediment Analyzer*
- Minimal consumables: Only cuvettes are needed – no liquid reagents with consumable exchanging taking less than 3 minutes
Clinical Use
Menu for the Atellica® 1500 Automated Urinalysis System*
CLINITEK Novus Analyzer® with CLINITEK Novus Cassettes | Atellica UAS 800 Analyzer* | |
| CLINITEK Novus 10 Cassette | CLINITEK Novus PRO 12 Cassette* | Auto-reported Categories |
| Bilirubin | Albumin | Bacteria |
| Clarity | Albumin-to-creatine ratio | Crystals |
| Color | Bilirubin | Hyaline cast |
| Glucose | Clarity | Mucus |
| Ketone | Color | Nonsquamous EC |
| Leukocyte esterase | Creatine | Pathalogical cast |
| Nitrite | Glucose | RBC |
| Occult blood | Ketone | Sperm |
| pH | Leukocyte esterase | Squamous EC |
| Protein | Nitrite | WBC |
| Specific gravity | Occult blood | Yeast |
| Urobilinogen | pH | Sub-categories: |
| Protein | Bacteria cocci | |
| Protein-to-creatine ratio | Bacteria rods | |
| Specific gravity Urobilinogen | WBC clumps | |
Technical Details
| Physical Information | |
| Depth (with rack handler) | 680 mm |
| Height (with rack handler) | 625 mm |
| Width (with rack handler) | 1260 mm |
| Weight | 114 kg |
| System Information | |
| Throughput | 106–240 samples/hr |
| Storage | 10,000 patient results including images |
| Decibel rating | < 60 dB |
| STAT function | Yes |
| Sediment Result units | cells/μl, cells/hpf, cells/lpf |
| Chemistry Result units | Conventional, Conventional with plus system, SI, and SI with plus system |
| Time to first result | < 4 minutes |
| Sample loading capacity | 100 samples |
| Interface | TCP/IP or Serial - ASTM and HL7 |
| Sediment Menu | |
| • RBC (red blood cell) • WBC (white blood cell) – WBCc (white blood cell clump) • BAC (bacteria) – BACr (rod bacteria) – BACc (coccus bacteria) • EPI (squamous epithelial cell) | • NEC (nonsquamous epithelial cell) • PAT (pathological cast) • HYA (hyaline cast) • CRY (crystal) • YEA (yeast) • MUC (mucus) • SPRM (sperm) |
| Chemistry Menu | |
| CLINITEK Novus Analyzer with CLINITEK Novus Cassettes | |
CLINITEK Novus 10 Cassette
| CLINITEK Novus PRO 12 Cassette† • Bilirubin • Clarity • Color • Glucose • Ketone • Leukocyte esterase • Nitrite • Occult blood • pH • Protein • Specific gravity • Urobilinogen • Albumin • Albumin-to-creatinine ratio • Creatinine • Protein-to-creatinine ratio |
| Electrical Requirements | |
| Main unit | 110–240 VAC | 50–60 Hz | 210 W max. Fuse rating: T8A 250 V |
| Personal computer with WINDOWS 7 operating system | 100–240 VAC | 50–60 Hz | 400 W max. |
| Monitor | 100–240 VAC | 50–60 Hz | 100 W max. |
| Environmental Specifications | |
| Operating temperature | 18–30°C (64–86°F) |
| Operating humidity | 20–80% RH, noncondensing |
| Altitude range of operation | 0–2000 m above sea level |
| Storage temperature | -25 to +60°C (-13 to +140°F) |
| Storage humidity | 20–80% RH , noncondensing |
| Storage atomospheric pressure | 106–70kPA (approximately 0–3000 m) |
| Compliance | TUV, CE, EMC, IEC 61326 |
| Supported Bar-code Specifications | |
| Supported bar-code types† | Codabar Code 39 Code 93 Code 128 Interleaved 2 of 5 |
| Test Rack and Tube Specifications | |
| Test tube racks supported | Only test tube racks provided by Siemens Healthineers |
| Minimum sample volume in tube | 2.6 mL |
| Height | 95–106 mm (3.74–4.17 inches) |
| Rack opening diameter | 16.5 mm (0.64 inches) |
| Cuvettes | |
| Supported types | Only Atellica UAS 800 cuvette cartridges supplied by Siemens Healthineers |
| Number of tests per cartridge | 600 cuvettes |
| Shelf life | 2 years |
| Test Cassettes | |
| Supported types | CLINITEK Novus 10 Cassette CLINITEK Novus PRO 12 Cassette* |
Share this page:
Related Products, Services & Resources
*Not available for sale in U.S. Product availability varies by country.
†External handheld bar-code reader optional
Atellica, CLINITEK Novus, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. All other trademarks and brands are the property of their respective owners.
The products/features (mentioned herein) are not commercially available in all countries. Due to regulatory reasons their future availability cannot be guaranteed. Please contact your local Siemens organization for further details.