A mobile cardiac myxoma

Liangping Ni, MD1; Huan Li, MD1; Longsheng Wang, MD1; Pengyun Cheng, MD2; Xinglong Liu, MD2

1 Department of Radiology, The Second Hospital of Anhui Medical University, Hefei, P. R. China

2 Siemens Healthineers, China

2021-05-28
A 74-year-old female patient, suffering from exertional chest tightness and shortness of breath for the past year, came to the hospital for a check up. Her symptoms could be relieved by rest, however, had worsened over the past two months. An echocardiography examination performed in the local hospital revealed an abnormal growth in the left atrium. She was referred to our hospital for treatment. A cardiac CT was requested for pre-operative evaluation.
CT images revealed normal coronary arteries without plaques or stenoses (Fig. 1). A large, polypoid mass was seen in the left atrium, attached to the interatrial septum. The mass was smoothly demarcated and homogeneously hypodense, without enhancement or calcification. It prolapsed partially into the left ventricle through the mitral valve in the diastole and completely returned into the left atrium in the systole (Fig. 2 and Fig. 3). The dynamic movement and the stalk of the mass were better visualized when using multi-phase image reconstructions and the cine mode display. A mobile cardiac myxoma was suspected and was confirmed through pathology after the surgical resection. The patient recovered uneventfully.
Curved MPR images and a cVRT image reveal normal coronary arteries.
Courtesy of Department of Radiology, The Second Hospital of Anhui Medical University, Hefei, P. R. China
Fig. 1: Curved MPR images (Figs. 1a-1c) and a cVRT image (Fig. 1d) reveal normal coronary arteries.
MPR images in long axis show a large, hypodense mass remaining in the left atrium in systole, and prolapsing through the mitral valve into the left ventricle in diastole.
Courtesy of Department of Radiology, The Second Hospital of Anhui Medical University, Hefei, P. R. China
Fig. 2: MPR (5 mm) images in long axis show a large, hypodense mass remaining in the left atrium in systole (Fig. 2a, arrow), and prolapsing through the mitral valve into the left ventricle in diastole (Fig, 2b, arrow).
Courtesy of Department of Radiology, The Second Hospital of Anhui Medical University, Hefei, P. R. China
Fig. 3: A cine display, using multi-phase image reconstructions, shows the dynamic movement of both the mass and the heart. It demonstrates that the mass partially prolapses into the left ventricle in diastole and completely returns into the left atrium in systole. The stalk is attached to the interatrial septum. The mitral valve opens and closes properly.
A cardiac myxoma (CM) is the most common benign primary tumor of the heart. It typically manifests as a polypoid intracardiac mass, most often in the left atrium, attached to the interatrial septum. The attachment point can be pedunculated, and a certain degree of mobility can occur depending upon the length of the stalk. The symptoms of the CM are usually unspecific or are similar to those of coronary artery disease (CAD). Surgical removal is considered the optimal treatment and must be carefully planned, especially when CM and CAD co-exist, and a combined operative strategy is contemplated. Therefore, it is important to evaluate the cardiac chambers as well as the coronary arteries prior to surgery. In this case, a cardiac CT is performed with prospective ECG triggering using one bolus of contrast injection. Multiple phases are reconstructed using acquired data to demonstrate the dynamic movement and the stalk of the mass. The best systolic and diastolic phases are automatically identified by the CT system. This significantly improves our daily workflow, with a great impact upon time-saving. The achieved image quality is optimal for a comprehensive evaluation, which enables the physicians to reach the diagnosis of CM, as well as to rule out CAD.

Scanner

Scan area

Heart

Scan mode

Prospective ECG triggered sequential scan

Scan length

113 mm

Scan direction

Cranio-caudal

Scan time

3.5 s

Tube voltage

90 kV

Effective mAs

227 mAs

Dose modulation

CARE Dose4D

CTDIvol

14.1 mGy

DLP

186.7 mGy*cm

Rotation time

0.25 s

Slice collimation

176 x 0.6 mm

Slice width

0.75 mm

Reconstruction increment

0.5 mm

Reconstruction kernel

Bv40

Heart rate

67 - 70 bpm

Contrast

320 mg/mL

Volume

60 mL + 40 mL saline

Flow rate

4.5 mL/s

Start delay

Bolus tracking with 100 HU at ascending aorta + 5 s