CLINITEK Atlas Automated Urine Chemistry Analyzer (Carousel)

Excellence in Automated Urinalysis - Productivity, Reliability, Flexibility, Ease of Use

The CLINITEK Atlas® Automated Urine Chemistry Analyzer is a fully automated, benchtop instrument designed for high-volume urine testing, using Siemens' proven technology in dry-pad urine chemistry analysis.

  • Unique, continuous reagent roll for simplified loading and reduced interruptions
  • Modular design for compatibility with laboratory automation systems
  • Built-in liquid level sensing for simplified sample preparation
Fale Conosco

Características e Benefícios

Liquid ready-to use calibration material ensures accuracy
Minimizes lot variability

Menu driven
Simple keypad operation

Adjusts automatically to sample volume
Pour any volume ≥ 2 mL without need for special steps or tubes
Reduces time to prepare samples for testing

Modular design facilitates track readiness

Sample loading options to meet testing and workflow needs. Loads trays with up to 50 samples or 40 samples plus calibrators and controls.
Trays available to accommodate different sample tube types

Assays

Format

Roll or dry chemistry strips affixed to plastic support

Maximum Tests On-board

490

On-board Stability

14 days

Tests Measured by Dry Pad

Leukocytes, Ketones, Protein, Glucose, Nitrite, Occult Blood, Urobilinogen, pH, Bilirubin, Color, Creatinine*, Protein-Low*

Tests by Physical Determinations

Specific Gravity by refractive index method; clarity by measurement of transmitted and scattered light

Dry Pad Test Measurements

Color change measured by reflectance photometry. Dual readings at reactive and reference wavelength

Especificações Técnicas

Dimensions

Depth: 64.8 cm (25.5 in)
Width: 72.4 cm (28.5 in)
Height:66.0 cm (26.0 in)

Weight
(instrument base only)

59.0 kg (130lb.)

Weight
(sample handler only)

7.2 kg (15.8 lb.)

Power Requirements

120 V 50/60 Hz, 100 V 50/60 Hz, 220-240 V 50/60 Hz

Ambient Operating Temperature Rate

18°C to 30°C (64°F to 86° F)

Ambient Operating Humidity Rate

20% to 80% relative humidity

Compliance

UL, CSA, CE

Throughput

1 sample every 16 seconds

Printer

External

Trays

Circular sample tray

Tray Capacity

50 samples (can be used as 40 = position tray dedicated spaces for calibrators, controls, STATs)

Sample Loading Capacity

1 tray up to 50 samples

Sample Tubes

Various depending on sample tray. Lipped tubes such as URIN-TEK or lipless up to 10.5 cm long

Sample Volume

Any volume from 2 mL to maximum fill of tube

Pipette

Precision impedance; ±1 µL accuracy

Rinse Cycle

Automatic pipette and SG well rinse between samples

Stat Sample Handling

Can be inserted during run; up to 2 samples

Bar Codes

Code 39, Interleaved 2 or 5, Codabar, Code 128

Memory

1000 patient records plus 200 calibration control records

Interfaces

ASTM or mode to emulate CLINITEK® 200+

Display

12 lines by 40 characters VGA display

Method

Wet calibration with 1 or 2 point calibration for all reagents

Frequency

With every Reagent Pak change (or every 24 hours when multiple sample-lot Reagent Paks used within 24 hours)